One cubic metre NIST traceable radon test chamber
1Rad Elec, Inc., 5714-C Industry Lane, Frederick, MD 21704, USA. pkotrappa@aol.com
Radiation Protection Dosimetry
|October 2, 2007
Summary
A new radon test chamber utilizes the NIST Radon Emanation Standard for calibrating detectors. This practical system accurately measures radon concentration, ensuring reliable radon monitoring equipment performance.
Area of Science:
- Environmental Science
- Radiological Health
- Metrology
Background:
- Accurate calibration of radon detectors is crucial for environmental monitoring and public health.
- Existing calibration methods may have limitations in terms of practicality and throughput.
- The National Institute of Standards and Technology (NIST) provides a Radon Emanation Standard for 226Ra.
Purpose of the Study:
- To develop and validate a practical radon test chamber for calibrating multiple radon detectors simultaneously.
- To establish a reliable method for generating known radon concentrations using the NIST standard.
- To assess the accuracy of the developed chamber by comparing calculated and measured radon levels.
Main Methods:
- A modified glove box was configured as a flow-through radon test chamber.
- Air was humidified and passed through the NIST Radon Emanation Standard (approx. 5000 Bq 226Ra).
- A theoretical expression was derived to calculate the radon concentration, including decay corrections for the system's volume and low flow rate.
Main Results:
- The radon test chamber successfully generated measurable radon concentrations.
- Calculated radon concentrations within the chamber showed good agreement with measurements from calibrated continuous radon monitors and passive integrating electret ion chambers.
- The system demonstrated suitability for calibrating multiple detectors concurrently.
Conclusions:
- The developed flow-through radon test chamber provides a practical and accurate method for detector calibration.
- The chamber leverages the NIST Radon Emanation Standard for reliable radon concentration generation.
- This system enhances the capacity for simultaneous calibration of various radon detection instruments.