Related Experiment Video
Updated: Jul 11, 2026

Quantifying Inferior Vena Cava Compliance and Distensibility in an In Vivo Ovine Model Using 3D Angiography
Published on: April 26, 2024
Loop-snare technique for difficult inferior vena cava filter retrievals
Lisa Rubenstein1, Albert K Chun, Megan Chew
1Department of Radiology, Division of Interventional Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.
Abstract:
Retrievable inferior vena cava (IVC) filters are generally retrieved without difficulties. However, when filters are tilted against the IVC wall, engaging the tip or hook of the filter can be difficult with the use of standard techniques. This report describes an alternative method of successful IVC filter retrieval by creating a wire loop between the filter legs, including the tip of the filter. This loop snare was successfully applied in eight cases after filter retrieval failed with the Recovery Cone or simple snare technique.