Detection of venous emboli using Doppler ultrasound
1Department of Vascular and Applied Surgery, Tokyo Medical and Dental University Graduate School, 1-5-45, Yushima, Bunkyo-ku, Tokyo 113-8519, Japan.
Objectives:
To detect emboli in the venous system using a Doppler ultrasound device with embolism detecting software.
Methods:
Pulmonary embolism (PE) was induced by injecting thrombus through the iliac vein of castrated swine under general anaesthesia. Data recorded from the Doppler system were analysed for high intensity transient signals (HITS) using receiver operating characteristic curves. Four different thrombi (5 or 10mm long and 3 or 5mm in diameter) were then injected to assess the quantitative analysis.
Results:
Thrombus could be detected in the venous system by the Doppler ultrasound device with an embolism detecting function. Appropriate confidence level was 60%. If thrombus were assumed to travel at the maximum flow rate (30 cm/s) in the inferior vena cava, the estimated embolism size was 10.4 S.D. 2.8mm for 3mm and 10.8 S.D. 4.9 mm for 5mm, both of which were close to 10mm.
Conclusion:
Thrombi could be detected as high intensity transient signals in the venous system. The appropriate confidence level was 60%. The size of emboli can be estimated if they are more than 3mm in diameter when the venous flow rate is 30 cm/s or less. Our results suggest that it may be possible to detect emboli in the subclavian vein, axillary vein or inferior vena cava in clinical cases.
Related Concept Videos
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Venous Thrombosis III: Interprofessional Care
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation