Technical report: optimal quantity of saline for epidural pressure waveform analysis
Etienne de Médicis1, Jocelyn Pelletier, René Martin
1Département d'anesthésiologie du Centre Hospitalier Universitaire de Sherbrooke, Sherbrooke, Québec, Canada. estria1@globetrotter.net
Canadian Journal of Anaesthesia = Journal Canadien D'Anesthesie
|October 16, 2007
Summary
Three-hole epidural catheters (S3) are superior to single-hole catheters (S1) for epidural waveform analysis (EPWA). Optimal saline volume for EPWA is 5 mL, ensuring accurate epidural catheter placement.
Area of Science:
- Anesthesiology
- Pain Management
- Medical Devices
Background:
- Epidural waveform analysis (EPWA) confirms optimal epidural catheter placement by assessing thecal pressure waveform transmission.
- The ideal injectate volume for optimizing EPWA remains underexplored.
Purpose of the Study:
- To determine the optimal injectate volume for epidural waveform analysis (EPWA).
- To compare the responsiveness of different epidural catheter types for EPWA.
Main Methods:
- A randomized controlled trial involving 60 patients compared three epidural catheter types: stimulating single-hole (S1), stimulating three-hole (S3), and non-stimulating three-hole (NS3).
- Epidural pressure waveform transmission was assessed using serial 2.5 mL saline injections up to 10 mL.
- Patient follow-up by acute pain services confirmed adequate postoperative epidural analgesia.
Main Results:
- Three-hole catheters (S3 and NS3) demonstrated significantly higher responsiveness compared to single-hole catheters (S1) at low injectate volumes (P = 0.003 for S3, P = 0.012 for NS3).
- No significant difference in waveform transmission was observed between S3 and NS3 catheters (P = 0.25).
- Optimal responsiveness was achieved with 5 mL of saline for S1 catheters and 2.5 mL for S3 catheters, with no further improvement beyond these volumes.
Conclusions:
- Stimulating three-hole (S3) epidural catheters are more effective than single-hole (S1) catheters for waveform transmission during EPWA at low injectate volumes.
- An initial injectate volume of 5 mL is recommended for EPWA, irrespective of catheter type, to ensure optimal results.