Ozone air levels adjacent to a dental ozone gas delivery system
Elisabeth Johansson1, Ingrid Andersson-Wenckert, Annika Hagenbjörk-Gustafsson
1Dental Hygienist Education, Department of Odontology, Umeå University, Umeå, Sweden. Elisabeth.Johansson@odaont.umu.se
Objective:
Ozone (O3) has been suggested as an anti-microbial treatment in dentistry, with an ozone gas delivery system introduced for the treatment of fissure and root caries. The aim of this study was to investigate the sealing capacity of the novel delivery system and its re-suction capacity during accidental displacement of the cup at different stages of ozone delivery.
Material And Methods:
Ozone leakage was studied in vitro after application on a flat metal surface and on buccal and occlusal tooth surfaces. An ozone analyzer was used to measure ozone gas concentrations adjacent to the delivering cups when adapted to the target surfaces during and after 10-20 s application cycles. The measured levels were compared with the background concentrations in the room. Measurements were performed 1) after complete ozone application cycles, 2) within the cycle before the start of the suction period, and 3) after displacements of the cup during the cycles.
Results:
Ozone air values varied between 8 and 166 microgram/meter-3 for the flat metal surface and between 0 and 108 microgram/meter-3 for the tooth surfaces. Ozone leakage levels were 7.6 microgram/m-3 for the flat and and 7.4 microgram/meter-3 and 5.6 microgram/meter-3 for the buccal and occlusal surfaces, respectively, and 5.2 microgram/meter-3 and 9.8 microgram/meter-3 for the premolar and molar surfaces, respectively. Cycles with displacement showed significantly higher leakage levels than continuous complete cycles (p=0.03).
Conclusions:
Ozone application cycles with displacements showed significantly higher leakage levels than continuous complete cycles. The largest ozone delivery cups showed the highest leakage values. A change in background levels was seen with similar change in adjacent ozone levels. The overall measured ozone leakage values were low after normally functioning delivery cycles and after repeated displacements. The delivery system can be considered safe.
More Related Videos
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Inhalational Anesthetics: Overview
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...