Related Experiment Video
Updated: Jul 10, 2026

A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Noninvasive central venous pressure measurement by controlled compression sonography at the forearm
Christoph Thalhammer1, Markus Aschwanden, Angela Odermatt
1Department of Vascular Medicine, University Hospital, Basel, Switzerland. thalhammerc@uhbs.ch
Objectives:
We sought to validate a new noninvasive technique to determine central venous pressure (CVP) using high-resolution compression sonography.
Background:
Information concerning CVP is crucial in clinical situations, including cardiac failure, volume overload, and sepsis. The measurement of CVP, however, requires puncture of a vein with attendant risk of complication.
Methods:
After a proof-of-concept study in healthy subjects, a prospective blinded evaluation was performed comparing CVP measurement using a central venous catheter with measurement using compression sonography in critically ill (intensive care unit) patients.
Results:
In healthy subjects with experimentally induced venous hypertension with a wide range of pressure values, a strong correlation (r = 0.95; p < 0.001) between noninvasive and invasive peripheral venous pressure at the forearm was shown. High interobserver agreement with an intraclass correlation coefficient of 0.988 shows excellent reliability of the system. Noninvasive peripheral venous pressure measurement at the forearm showed a good correlation with CVP in 50 intensive care unit patients with the forearm positioned both below heart level (r = 0.84; p < 0.001) and at heart level (r = 0.85; p < 0.001). The mean difference between invasive and noninvasive measurement was negligible (-0.1 +/- 3.5 cm H2O and -0.7 +/- 3.4 cm H2O, respectively).
Conclusions:
Controlled-compression sonography is a valuable tool for measuring venous pressure in peripheral veins and allows reliable indirect assessment of CVP without intravenous catheterization.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...