[Electromagnetic interference of electrical dental equipment with cardiac pacemakers]
H S Brand1, E V van der Hoeff, T A M Schrama
1Uit de afdeling Tandheelkundige Basiswetenschappen, sectie Orale Biochemie, en de afdeling Mondziekten en Kaakchirurgie van het Academisch Centrum Tandheelkunde Amsterdam (ACTA). hs.brand@vumc.nl
Abstract:
Eight different electrical dental appliances were tested at different intervals for their ability to interfere with the function of a contemporary cardiac pacemaker. The normal atrial and ventricular pacing was inhibited by an ultrasonic bath cleaner at a distance of less than 15 cm. In contrast, a dental chair, an electrosurgical unit, an ultrasonic tooth scaler, 2 handpieces, and 2 amalgamators failed to produce electromagnetic interference at the minimum distance of 2.5 cm. In conclusion, the results suggest that normal clinical use of dental electrical equipment does not have any significant effect on the cardiac pacemaker tested.
Related Concept Videos
Mechanism of Cardiac Arrhythmias
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Electrophysiology of Normal Cardiac Rhythm
Imaging Studies for Cardiovascular System IV: CMRI
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...