Biocompatibility of candidate materials for the realization of medical microdevices
Pierre Pouponneau1, L'Hocine Yahia, Yahye Merhi
1Laboratory for the Innovation and Analysis of the Bioperformance, Ecole Polytechnique de Montreal, Department of Computer Engineering, Institute of Biomedical Engineering, Québec, Canada. pierre.pouponneau@polymtl.ca
Abstract:
The propulsion of ferromagnetic micro-carriers in the blood vessels by magnetic gradients generated from a Magnetic Resonance Imaging (MRI) system is of special interest for targeted interventions such as chemotherapy or chemo-embolization. As such, Fe-Co alloys for its highest magnetization saturation, and single crystal Ni-Mn-Ga powder and Terfenol-D for their deformation in magnetic field are evaluated for their biocompatibility. The toxicity of these materials is evaluated with MTT cell viability tests. The tests show that Fe-Co (Permendur and Vacoflux 17) alloys are toxic within 24 hours while the single crystal Ni-Mn-Ga powder becomes toxic after 48 hours. The Terfenol-D, despite its high degradation, has 90% cell viability after 72 hours. These results indicate that such candidate materials to be considered in untethered micro-carriers or devices in the blood vessels would require, depending upon the time spent in the blood vessels, further processes to be viable for such applications.