Cardiac Pressure-Volume Loop Analysis Using Conductance Catheters in Mice
Published on: September 17, 2015
The effect of fiber orientation on volume measurement using conductance catheter techniques
1Sch. of Eng., Tasmania Univ., Hobart, Tasmania, Australia. ct0@postoffice.utas.edu.au
Abstract:
Estimation of parallel conductance using the impedance electrode technique is usually done assuming isotropic conditions. This may not be the best solution since the myocardium is an anisotropic material. This paper exposes the effect of fiber orientation for volume measurement using a conductor model with asymmetrical source electrodes. Simulation results show calculated volumes between surrounding materials with and without myocardial fiber orientation included in the model. We plan to extend these study results to the real heart for developing conductance catheter techniques for use in blood volume measurements in the right ventricle.