Automated, High-resolution Mobile Collection System for the Nitrogen Isotopic Analysis of NOx
Published on: December 20, 2016
On-line technique to determine the isotopic composition of total dissolved nitrogen
Dries Huygens1, Pascal Boeckx, Jan Vermeulen
1Laboratory of Applied Physical Chemistry ISOFYS, Faculty of Bioscience Engineering, Ghent University, Ghent, Belgium. Dries.Huygens@UGent.be
Abstract:
A new on-line analytical setup for 15N measurements of total dissolved nitrogen (TDN) has been developed through the coupling of a high-temperature catalytic (Ce(IV)O2) oxidation furnace, a Cu reduction furnace, and an isotope ratio mass spectrometer. The detection limit for accurate delta15N measurements is 20 mg of N L-1. For N-containing compounds dissolved in water, a standard deviation on N concentration measurements of 0.2 mg of N L-1, independent of N concentration, has been found. Reproducibility on delta15N increased with increasing N concentration, with standard deviations varying from 0.8 to 0.1 per thousand in the range of 20-100 mg of N L-1. Salt matrixes, in which the N species might be dissolved, could influence the analysis capacity and continuity, mainly at concentrations above 0.1 M. To our knowledge, this system is the first successful on-line setup capable of performing routine delta15N and N concentration measurements of the TDN pool. Potential applications are large and are believed to result in an increased insight in N cycling and dissolved organic nitrogen behavior in terrestrial and aquatic ecosystems.