Bacterial endospore inactivation caused by outgassing of vapourous hydrogen peroxide from polymethyl methacrylate
P A Baron1, C F Estill, J K Beard
1National Institute for Occupational Safety and Health, Centers for Disease Control and Prevention, 4676 Columbia Parkway, Cincinnati, OH 45226, USA. pbaron@cdc.gov
Aims:
To investigate the cause and to eliminate the inactivation of Bacillus anthracis strain Sterne spores settled onto agar and stainless steel surfaces in plastic holders.
Methods And Results:
In an experimental chamber in which spores settled onto sampling surfaces, vapourous hydrogen peroxide (VHP) was used for decontamination between experiments. It was demonstrated that hydrogen peroxide (H(2)O(2)) absorbed into plastic (Plexiglas) surfaces and could outgas in the sample holders. Further experiments demonstrated that H(2)O(2) was released from Plexiglas sample holders in sufficient quantity to inactivate spores. High temperature degassing (30-35 degrees C) for several days or aluminum coating of the surfaces were two remedies found to be effective in preventing inadvertent spore inactivation.
Conclusions:
H(2)O(2) can be absorbed into plastic and released after an extended period of time (weeks), allowing a sufficient concentration to accumulate in small volumes to inactivate spores. Outgassing the plastic or coating the surface with an impermeable layer are potential solutions to reduce spore inactivation.
Significance And Impact Of The Study:
Many studies with bacilli and other organisms are carried out using small plastic containers that may have been sterilized using H(2)O(2) or other agents. This study presents a cautionary note to ensure elimination of H(2)O(2) or other sterilizing agents to prevent spurious results.
Related Concept Videos
Endospores and Sporulation
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Physical Methods for Controlling Microbial Growth: Temperature
Gene Regulation During Sporulation
Physical Methods for Controlling Microbial Growth: Radiation and Filtration