Minimizing RF heating of conducting wires in MRI
Christopher J Yeung1, Parag Karmarkar, Elliot R McVeigh
1Laboratory of Cardiac Energetics, National Heart, Lung & Blood Institute, Division of Intramural Research, National Institutes of Health, Bethesda, Maryland 20892-1061, USA.
Abstract:
Performing interventions using long conducting wires in MRI introduces the risk of focal RF heating at the wire tip. Comprehensive EM simulations are combined with carefully measured experimental data to show that method-of-moments EM field modeling coupled with heat transfer modeling can adequately predict RF heating with wires partially inserted into the patient-mimicking phantom. The effects of total wire length, inserted length, wire position in the phantom, phantom position in the scanner, and phantom size are examined. Increasing phantom size can shift a wire's length of maximum tip heating from about a half wave toward a quarter wave. In any event, with wires parallel to the scanner bore, wire tip heating is minimized by keeping the patient and wires as close as possible to the central axis of the scanner bore. At 1.5T, heating is minimized if bare wires are shorter than 0.6 m or between approximately 2.4 m and approximately 3.0 m. Heating is further minimized if wire insertion into phantoms equivalent to most aqueous soft tissues is less than 13 cm or greater than 40 cm (longer for fatty tissues, bone, and lung). The methods demonstrated can be used to estimate the absolute amount of heating in order to set RF power safety thresholds.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Field Due To A Thin Straight Wire
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Field Due to Two Straight Wires