Determination of alcohol compounds using corona discharge ion mobility spectrometry
Hai-yan Han1, Guo-dong Huang, Shun-ping Jin
1Laboratory of Environmental Spectroscopy and Key Laboratory of Environmental Optics and Technology, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China. hanhy0226@yahoo.com.cn
Abstract:
Ion mobility spectrometry (IMS) is a very fast, highly sensitive, and inexpensive technique, it permits efficient monitoring of volatile organic compounds like alcohols. In this article, positive ion mobility spectra for six alcohol organic compounds have been systematically studied for the first time using a high-resolution IMS apparatus equipped with a discharge ionization source. Utilizing protonated water cluster ions (H2O)n H+ as the reactant ions and clean air as the drift gas, alcohol organic compounds, ethanol, 1-propanol, 2-propanol, 1-butanol, 1-pentanol and 2-octanol, all exhibit product ion characteristic peaks in their respective ion mobility spectrometry, that is a result of proton transfer reactions between the alcohols and reaction ions (H2O)n H+. The mixture of these alcohols, including two isomers, has been detected, and the results showed that they could be distinguished effectively in the ion mobility spectrum. The reduced mobility values have been determined, which are in very well agreement with the traditional 63Ni-IMS experimental values. The exponential dilution method was used to calibrate the alcohol concentrations, and a detection limit available for the alcohols is in order of magnitude of a few ng/L.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
08:23Analyzing the Photo-oxidation of 2-propanol at Indoor Air Level Concentrations Using Field Asymmetric Ion Mobility Spectrometry
Published on: June 14, 2018
Related Concept Videos
Mass Spectrometry: Alcohol Fragmentation
Atomic Absorption Spectroscopy: Atomization Methods
Ion-Exchange Chromatography
IR and UV–Vis Spectroscopy of Aldehydes and Ketones
Chemical Ionization (CI) Mass Spectrometry
IR and UV–Vis Spectroscopy of Carboxylic Acids
However, the stretching absorptions for the C=O bond vary depending on the structure of carboxylic acids. The C=O bond of the free carboxylic acids shows a higher stretching frequency, 1760 cm−1, while H-bonded carboxylic acids (dimers) exhibit stretching absorptions at a lower frequency, 1710 cm−1. The C=O bond of the...