Influence of intermittent pneumatic compression devices on non-invasive blood pressure measurement of the ankle
Related Experiment Video
Updated: Jul 10, 2026

Computerized Dynamic Posturography for Postural Control Assessment in Patients with Intermittent Claudication
Published on: December 11, 2013
Bijal R Parikh1, Abraham M Simon, John N Kouvaras
1Department of Anesthesiology, Saint Barnabas Medical Center, Livingston, NJ 07039, USA. bijalparikh@hotmail.com
Background:
It is rapidly becoming the standard of care to employ the perioperative use of intermittent pneumatic compression (IPC) devices on the lower leg to prevent the development of deep venous thrombosis. There are instances when the anesthesiologist must apply a blood pressure cuff to the ankle in order to obtain a blood pressure reading. The purpose of this study is to determine the influence of IPC devices on non-invasive blood pressure measurements of the ankle.
Methods:
We utilized the Flowtron Excel External Pneumatic Compression System Model AC550 and the Datascope Passport non-invasive blood pressure monitor Model EL, in 50 healthy volunteers ranging in ages from 23 to 35 who were free of systemic disease. The IPC devices were applied to both lower extremities and sequentially inflated. All blood pressure measurements (systolic, diastolic, and mean arterial pressure) were taken with the volunteer in the supine position, obtained in triplicate and then averaged. Blood pressure measurements were obtained at four time periods: 1) when the IPC device was placed over the ankle blood pressure cuff but not inflated (control) 2) 15 s prior to inflation of the IPC device 3) at inflation of the IPC device and 4) 15 s after the beginning of inflation of the IPC device. The difference between the averaged blood pressures and the control blood pressures were calculated. The resulting values were then converted to absolute values. The averaged absolute values were analyzed using ANOVA for repeated (paired) data.
Results:
No statistical significant difference was shown between the control blood pressure measurements and any of the three sets of blood pressure measurements relative to IPC device inflation; 15 s prior, at onset, and 15 s after inflation. The p-values for systolic, diastolic, and MAP were respectively 0.872, 0.645 and 0.522. With these p-values, the power tests for each were >0.80.
Conclusion:
The IPC device had no significant impact on non-invasive blood pressure measurement of the ankle in our study.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure in the Leg
Preparation:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Assessment of blood pressure in brachial artery(two-step method)