Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
A continuous and non-invasive arterial pressure monitoring system in dialysis patients
Emanuele Mambelli1, Elena Mancini, Antonio Santoro
1Nephrology, Dialysis and Hypertension Unit, Policlinico S. Orsola-Malpighi, Bologna, Italy.
Background:
As symptomatic intradialytic hypotension in the hemodialysis (HD) patient is often a sudden event whose onset cannot be predicted by means of extemporary measures, continuous blood pressure (BP) measurement would be far more useful. We tested a new continuous noninvasive BP monitoring system, Harmonized Alert Sensing Technology (HASTE), which, by means of the analysis of the finger pulse wave, obtained from an O2 sensor, estimates a beat-to-beat systolic pressure value (Esys) and supplies a continuous read-out. The study aim sought to verify the reliability of this non-invasive instrument in continuously providing systolic pressure values during HD.
Methods:
We studied 18 patients during HD treatment, initially comparing the Esys with the invasive blood pressure measurement (SYS). Subsequently, the Esys derived from the O2 sensor in the arm with the shunt (S) and the arm without (N), respectively, were both compared with the cuff measurement.
Results:
The mean difference between SYS and Esys was 0.7 +/- 13.3 mm Hg (p < 0.01; r = 0.80). There was a mean difference of 0.2 +/- 21.9 mm Hg (p = NS; r = 0.67) between Esys(N) and Esys(S). The correlation was not statistically significant even between Esys(N) or Esys(S) versus the cuff measurement, respectively. Patient movement compromised the accuracy of the estimations made using the O2 sensor in the non-fistula arm.
Conclusions:
A good correlation between the data estimated by HASTE compared with invasive BP suggests that the instrument may prove useful for continuously monitoring the blood pressure trends during the dynamic hemodialysis situation. However, its sensibility needs to be improved in order to be used indifferently in both arms with a view to achieving real intradialytic hypotension prevention.
Related Concept Videos
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Continuous Renal Replacement Therapy
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Hemodialysis I: Introduction
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: