Mass and position determination of attached particles on cantilever based mass sensors
1MIC, Department of Micro- and Nanotechnology, NanoDTU, Technical University of Denmark, Building 345E, DK-2800 Lyngby, Denmark. sd@mic.dtu.dk
Abstract:
An analytical expression relating mass and position of a particle attached on a cantilever to the resulting change in cantilever resonant frequency is derived. Theoretically, the position and mass of the attached particle can be deduced by combining measured resonant frequencies of several bending modes. This finding is verified experimentally using a microscale cantilever with and without an attached gold bead. The resonant frequencies of several bending modes are measured as a function of the bead position. The bead mass and position calculated from the measured resonant frequencies are in good agreement with the expected mass and the position measured.
Related Concept Videos
Mass Analyzers: Overview
Mass Analyzers: Common Types
Moments and Centers of Mass
Mass Spectrometers
Significance of Center of Mass
Rigid Body Equilibrium Problems - II
Consider two children sitting on a seesaw, which has negligible mass. The first child has a mass (m1) of 26 kg and sits at point A, which is 1.6 meters (r1) from the pivot point B; the second child has a mass (m2) of 32 kg and sits at point C. How far from the pivot point B should the second child sit (r2) to balance the seesaw?