Diagnostic accuracy of dual-source computed tomography in the diagnosis of coronary artery disease
Thorsten R C Johnson1, Konstantin Nikolaou, Stephanie Busch
1Department Clinical Radiology, Medical Clinic I, University of Munich, Grosshadern Campus, Munich, Germany. thorsten.johnson@med.uni-muenchen.de
Insights
Dual-source computed tomography (DSCT) demonstrates high accuracy in diagnosing coronary artery disease (CAD), with excellent sensitivity and negative predictive value for ruling out significant CAD per patient.
Area of Science:
- Cardiovascular Imaging
- Radiology
- Medical Diagnostics
Background:
- Coronary artery disease (CAD) diagnosis relies on invasive methods.
- Non-invasive imaging modalities are sought for accurate CAD assessment.
Purpose of the Study:
- To evaluate the diagnostic accuracy of dual-source computed tomography (DSCT) compared to invasive coronary angiography.
- Assess DSCT performance on both per-patient and per-segment bases for CAD detection.
Main Methods:
- 35 patients with suspected CAD underwent both DSCT and quantitative x-ray coronary angiography (QCA).
- DSCT parameters included 83 ms temporal resolution and ECG-gated modulation.
- Sensitivity, specificity, PPV, and NPV were calculated for both methods.
Main Results:
- DSCT achieved 100% sensitivity and 89% specificity per patient for significant CAD (>50% stenosis).
- Per-segment analysis showed 88% sensitivity and 98% specificity.
- DSCT demonstrated high negative predictive value (100% per patient, 99% per segment).
Conclusions:
- DSCT offers robust image quality and high diagnostic accuracy for CAD detection.
- Maximal sensitivity and NPV in per-patient analysis highlight DSCT's strength in excluding significant CAD.
Purpose:
The aim of this study was to evaluate the diagnostic accuracy of dual-source computed tomography (DSCT) with reference to invasive coronary angiography in the diagnosis of coronary artery disease (CAD) on a per-patient as well as on a per-segment basis.
Materials And Methods:
Thirty-five patients with known or suspected CAD underwent both DSCT (Somatom Definition, Siemens Medical Solutions) and quantitative x-ray coronary angiography (QCA). Parameters of CT acquisition were gantry rotation time 0.330 seconds (ie, temporal resolution 83 milliseconds), tube voltage 120 kV, tube current 560 mA with ECG-triggered tube current modulation and full current at 70% of the cardiac cycle for heart rates below 70 beats per minute or full current between 30% and 80% for higher and arrhythmic heart rates. The pitch was also adapted to the heart rate, ranging from 0.2 to 0.43. Volume and flow rate of contrast material (Ultravist 370, Schering AG) were adapted to the patient's body weight. Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of DSCT in the detection or exclusion of significant CAD (ie, stenoses >50%) were evaluated on a per-patient and per-segment basis.
Results:
All 35 CT angiograms were of diagnostic image quality. QCA demonstrated significant CAD in 48% (n = 17) and nonsignificant disease or normal coronary angiograms in 52% (n = 18) of the patients. Sensitivity, specificity, PPV, and NPV of DSCT on a per-patient basis were 100%, 89%, 89%, and 100%, respectively. On a per-segment basis, 473 of 481 coronary artery segments were assessable (98%). QCA demonstrated stenoses >50% in 32 segments (7%), and no disease or nonsignificant disease in 433 segments (93%). For the detection of stenoses >50% on a per-segment basis, DSCT showed a sensitivity, specificity, PPV, and NPV of 88%, 98%, 78%, and 99%, respectively.
Conclusions:
The comparison of coronary DSCT with QCA shows a very robust image quality and a high diagnostic accuracy in a patient-based as well as a per-segment analysis. Maximal sensitivity and NPV in the per-patient analysis show the strength of the technique in ruling out significant CAD.
More Related Videos
06:57Semi-Automatic Graphical Tool for Measuring Coronary Artery Spatially Weighted Calcium Score from Gated Cardiac Computed Tomography Images
Published on: September 22, 2023
06:16Signal Acquisition, Score Interpretation, and Economics of a Non-Invasive Point-of-Care Test for Coronary Artery Disease
Published on: August 9, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Acute Coronary Syndrome III: Diagnostic Studies
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography