Optimization of image acquisition techniques for dual-energy imaging of the chest
N A Shkumat1, J H Siewerdsen, A C Dhanantwari
1Department of Medical Biophysics, University of Toronto, Toronto, Ontario, Canada M5G 2M9.
Abstract:
Experimental and theoretical studies were conducted to determine optimal acquisition techniques for a prototype dual-energy (DE) chest imaging system. Technique factors investigated included the selection of added x-ray filtration, kVp pair, and the allocation of dose between low- and high-energy projections, with total dose equal to or less than that of a conventional chest radiograph. Optima were computed to maximize lung nodule detectability as characterized by the signal-difference-to-noise ratio (SDNR) in DE chest images. Optimal beam filtration was determined by cascaded systems analysis of DE image SDNR for filter selections across the periodic table (Z(filter) = 1-92), demonstrating the importance of differential filtration between low- and high-kVp projections and suggesting optimal high-kVp filters in the range Z(filter) = 25-50. For example, added filtration of approximately 2.1 mm Cu, approximately 1.2 mm Zr, approximately 0.7 mm Mo, and approximately 0.6 mm Ag to the high-kVp beam provided optimal (and nearly equivalent) soft-tissue SDNR. Optimal kVp pair and dose allocation were investigated using a chest phantom presenting simulated lung nodules and ribs for thin, average, and thick body habitus. Low- and high-energy techniques ranged from 60-90 kVp and 120-150 kVp, respectively, with peak soft-tissue SDNR achieved at [60/120] kVp for all patient thicknesses and all levels of imaging dose. A strong dependence on the kVp of the low-energy projection was observed. Optimal allocation of dose between low- and high-energy projections was such that approximately 30% of the total dose was delivered by the low-kVp projection, exhibiting a fairly weak dependence on kVp pair and dose. The results have guided the implementation of a prototype DE imaging system for imaging trials in early-stage lung nodule detection and diagnosis.
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...