Related Experiment Video
Updated: Jul 7, 2026

Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
Published on: October 20, 2016
Real-time three-dimensional echocardiography in aortic stenosis: a novel, simple, and reliable method to improve
Juan Luis Gutiérrez-Chico1, José Luis Zamorano, Elsa Prieto-Moriche
1Unidad de Cardiología Intervencionista, Hospital de Meixoeiro, Instituto Galego de Medicina Técnica, Crtra. de Meixoeiro s/n, 36204 Vigo (Pontevedra), Spain. jlgutierrez@medynet.com <jlgutierrez@medynet.com>
Aims:
The aim of the study was to validate a novel formula for aortic area, based on the principle of continuity equation (CE), that substitutes Doppler-derived stroke volume (SV) by SV directly measured with real-time three-dimensional (RT3D) echo and semi-automated border detection. RT3D has proved outstanding accuracy for left ventricular volume calculation. So far, however, neither this potential has been applied to haemodynamic assessment, nor RT3D has succeeded in the evaluation of aortic valve disease.
Methods And Results:
Aortic area was measured in 41 patients with aortic stenosis using Gorlin's equation, Hakki's formula, Doppler CE, two-dimensional Simpson's volumetric method, and by the novel RT3D method. RT3D has the best linear association and absolute agreement with Gorlin of all non-invasive methods r = 0.902, intraclass correlation coefficient (ICC) = 0.846, better than CE (r = 0.646, ICC = 0.626) and two-dimensional volumetric method (r = 0.627, ICC = 0.378). Linear and Passing-Bablok regression show that RT3D fits better to Gorlin (r(2) = 0.814) than CE (r(2) = 0.417) and two-dimensional method (r(2) = 0.393). Its accuracy is comparable to Hakki's formula, routinely employed in catheter laboratories. Inter- and intraobserver agreements (ICC) were, respectively, 0.732 and 0.985, better than CE (0.662, 0.857). RT3D also grades most efficiently the severity of aortic stenosis as mild, moderate, or severe (weighted kappa = 0.932). RT3D underestimates aortic area (95% CI 0.084-0.193). ROC curves, however, show that the optimal cutoff point to consider aortic stenosis severity remains close to 1 cm(2) (1.06 cm(2)).
Conclusions:
RT3D is more accurate than CE and than two-dimensional volumetric methods to calculate area and to grade the severity of aortic stenosis. Area obtained by three-dimensional echo is slightly underestimated, but its range is clinically negligible.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...