Related Experiment Video
Updated: Jul 10, 2026

Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
Quantitative measurement of CyberKnife robotic arm steering
K H Wong1, S Dieterich, J Tang
1Department of Radiology, Georgetown University, 2115 Wisconsin Avenue NW #603, Washington, DC 20007, USA. wong@isis.georgetown.edu
Abstract:
Respiratory motion is a significant and challenging problem for radiation medicine. Without adequate compensation for respiratory motion, it is impossible to deliver highly conformal doses to tumors in the thorax and abdomen. The CyberKnife frameless stereotactic radiosurgery system with Synchrony provides respiratory motion adaptation by monitoring skin motion and dynamically steering the beam to follow the moving tumor. This study quantitatively evaluated this beam steering technology using optical tracking of both the linear accelerator and a ball-cube target. Respiratory motion of the target was simulated using a robotic motion platform and movement patterns recorded from previous CyberKnife patients. Our results show that Synchrony respiratory tracking can achieve sub-millimeter precision when following a moving object.