Related Experiment Video
Updated: Jul 10, 2026

Quantification of Vascular Parameters in Whole Mount Retinas of Mice with Non-Proliferative and Proliferative Retinopathies
Published on: March 12, 2022
Accuracy of ROPtool vs individual examiners in assessing retinal vascular tortuosity
David K Wallace1, Sharon F Freedman, Zheen Zhao
1Department of Ophthalmology, Duke University School of Medicine, Durham, North Carolina USA. david.wallace@duke.edu
Objective:
To prospectively determine if tortuosity assessment by a computer program (ROPtool) that traces retinal blood vessels and measures their tortuosity was more accurate than that of individual pediatric ophthalmologists.
Methods:
One hundred eighty-five high-quality RetCam images from premature infants were circulated to 3 retinopathy of prematurity (ROP) experts and 3 other pediatric ophthalmologists ("examiners") who graded the tortuosity in each quadrant as normal, pre-plus, or plus. These same images were analyzed using ROPtool.
Results:
Using expert consensus as the standard, ROPtool's overall accuracy of 95% (175 of 185) for identifying tortuosity sufficient for plus disease was similar to that of examiner 1 (93%; 172 of 185; P = .50), examiner 2 (93%; 172 of 185; P = .50), and examiner 3 (91%; 168 of 185; P = .10). ROPtool's sensitivity of 97% (36 of 37) compared favorably with that of examiner 1 (65%; 24 of 37; P < .001), examiner 2 (70%; 26 of 37; P < .001), and examiner 3 (81%; 30 of 37; P = .06).
Conclusion:
Computer-assisted analysis of retinal images can potentially reduce subjectivity in the diagnosis of plus disease and optimize timing of follow-up and treatment for ROP.