Preliminary study of motion artifact rejection for NIBP measurement in an ambulance
Yoonseo Koo1, Jaemin Kang, Il Hyung Shin
1Interdisciplinary Program, Biomedical Engineering Major, Graduate School, Seoul National University, Seoul, Korea. kattack9@melab.snu.ac.kr
Abstract:
Motion artifact resulting from patient's movement is a significant source for disturbing accurate noninvasive blood pressure (NIBP) measurement. In an ambulance, patients are exposed to unstable circumstances due to vehicle's movement and vibration during emergency transportation. Since NIBP is indirectly measured using oscillometry based on the change of cuff pressure, it can be affected by motion artifact much more than other biosignals. In this paper, we developed a new NIBP system with improved accuracy by measuring acceleration of the system caused by patient's motion. The NIBP module including a 3-axis accelerometer was directly mounted on a cuff to minimize the interference induced through connecting tube. The results show that the proposed NIBP system has the capability to reject the interference of motion artifact effectively in an ambulance.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement