Related Experiment Video
Updated: Jul 10, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Evaluation of different meshing algorithms in the computation of defibrillation thresholds in children
Jeroen G Stinstra1, Matthew Jolley, Michael Callahan
1Scientific Computing and Imaging Institute, Cardio Vascular Research and Training Institute, University of Utah, Salt Lake City, UT, USA. jeroen@cvrti.utah.edu
Abstract:
In this paper we evaluate different meshing schemes to solve for the bioelectric fields that arise in the human body due to the defibrillation shock generated by an Implantable Cardiac Defibrillator, with particular emphasis on implantation in children. For children, the question of relative performance of different electrode locations remains open. Computational simulation is a critical tool to address this question, and mesh design is a critical component of such simulations. We use the SCIRun software package to address this simulation problem because it combines the powerful numeric tools required with interactive flexibility allowing easy comparison of both algorithms and electrode orientation. We describe a pipeline that starts with segmented CT-images and produces clinically useful parameters. Using this framework we report below that a meshing scheme using regularly spaced hexahedral elements which are locally refined around the electrodes constitute a quick and relatively accurate way of solving this problem.
Related Concept Videos
Mesh Analysis
A fundamental concept in mesh analysis is the definition of meshes and mesh currents. A mesh is a closed...
Cardiopulmonary Resuscitation III: AED Use