Related Experiment Video
Updated: Jul 10, 2026

Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere
Published on: April 30, 2018
Noninvasive thermometry in a reentrant resonant cavity applicator
Yasutoshi Ishihara1, Yuichiro Endo, Hiroshi Ohwada
1Department of Electrical Engineering, Nagaoka University of Technology, Nagaoka, Niigata 940-2188, Japan. ishihara@vos.nagaokaut.ac.jp
A new noninvasive thermometry method uses electrical field phase changes in a resonant cavity to monitor hyperthermia treatment temperature changes. This technique offers a simpler alternative to MRI for internal body temperature monitoring.
Area of Science:
- Biomedical Engineering
- Medical Physics
- Electromagnetics
Background:
- Hyperthermia treatment effectiveness is limited by accurate temperature monitoring.
- Current noninvasive temperature monitoring methods, like MRI, often require large, complex equipment.
Purpose of the Study:
- To propose a novel, noninvasive thermometry method for improved hyperthermia treatment.
- To develop a technique for reconstructing internal temperature distributions.
Main Methods:
- Utilizing the phase change of an electrical field within a resonant cavity, which correlates with temperature variations.
- Collecting line-integrated phase change data by rotating a cavity resonator.
- Reconstructing the temperature distribution using a back-projection algorithm and finite difference time domain (FDTD) numerical analysis.
Main Results:
- Demonstrated a method for noninvasively measuring temperature changes inside a resonant cavity.
- Numerical analysis using FDTD indicated the feasibility of the proposed thermometric approach.
- The method shows potential for monitoring temperature changes within the body without large setups.
Conclusions:
- The proposed noninvasive thermometry method, based on electrical field phase changes, is feasible for monitoring temperature during hyperthermia.
- This technique offers a potentially simpler and more accessible alternative to MRI for internal temperature monitoring.
More Related Videos
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Thermosensation
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...