Comparison of birdcage and phase array coil using FDTD for the B(1) homogeneity in high field MRI
1Department of Biomedical Engineering, Kyung Hee University, Yongin, Gyeonggi, Republic of Korea.
Abstract:
RF coils play an important role to acquire MR images with the maintenance of high homogeneity in high field MR system more than 3.0 T. Many kinds of RF coils such as birdcage coil, STR, surface coil, and phase array coil have been used, however, the good uniformity of a coil has always been an issue. In this paper, comparison of B(1) homogeneity between birdcage and phase array coil was investigated using FDTD method at 3.0 T MRI in order to develop RF coils with the high uniformity. Three different configurations of the FDTD simulation were performed like as using a free space configuration, water phantom configuration, and head mesh model. B1 homogeneity was calculated to the case of birdcage coil and 8-channel phase array coil in each configuration of simulation. Improvement on the homogeneity of the images and reduction of standing wave effect was achieved with comparing the real MR images with the result from simulation.