In vivo stimulation on rabbit retina using CMOS LSI-based multi-chip flexible stimulator for retinal prosthesis
T Tokuda1, R Asano, S Sugitani
1Nara Institute Science and Technology, Takayama-cho 8916-5, Ikoma, Nara, 630-0192, JAPAN. tokuda@ms.naist.jp
Abstract:
We have performed in vivo electric stimulation experiments on rabbit retina to demonstrate feasibility of CMOS LSI-based multi-chip flexible neural stimulator for retinal prosthesis. We have developed new packaging structure with an improved flexibility and device control system which totally controls the LSI-based multi-chip stimulator, counter electrode, and stimulation generator. We have implanted the fabricated multi-chip stimulator into sclera pocket for STS (Suprachoroidal Transretinal Stimulation) configuration. We successfully obtained EEP (Electrically Evoked Potential) on visual cortex evoked by the multi-chip stimulator.