Related Experiment Video
Updated: Jul 10, 2026

Construction of a Wireless-Enabled Endoscopically Implantable Sensor for pH Monitoring with Zero-Bias Schottky Diode-based Receiver
Published on: August 27, 2021
A dual band wireless power and FSK data telemetry for biomedical implants
L H Jung1, P Byrnes-Preston, R Hessler
1School of Biomedical Engineering, University of New South Wales, Sydney NSW 2052, Australia.
Abstract:
A dual-link coil arrangement and a novel digital frequency-shift keying (FSK) demodulator are presented. The primary application of this system is for inductively powered biomedical implants. The implant is provided with data and power via two separate links. Two sets of coils are used in an arrangement such that the magnetic interference between the two pairs is minimized. The demodulator circuitry presented relies solely on delaying elements, utilizing a delayed digital FSK signal to sample the original digital FSK signal. A synchronized clock can be derived from the FSK signals alone, however, by utilizing the power signal to obtain a synchronized clock, a higher data rate and a decrease in complexity of the receiver circuitry can be achieved. The system was implemented on the bench and experimentally tested at a data rate of 2.083 Mbps with zero bit error rate while receiving a 4.17/6.25 MHz FSK carrier signal synchronized with 2.083 MHz clock derived from the power carrier. The power link was set to provide 58mW.