Related Experiment Video
Updated: Jul 10, 2026

Detection of Targetable Alterations in Non-small Cell Lung Cancer using Next-generation Sequencing
Published on: October 10, 2025
Cost assessment of the automated VERSANT 440 Molecular System versus the semi-automated System 340 bDNA Analyzer
Tarek Elbeik1, Richard A Loftus, Scott Beringer
1Professional Habitat Design, LLC; Health Management Institute, Inc., c/o 584 Castro Street 349, San Francisco, CA 94114, USA. t.elbeik@phdllc.net
Abstract:
Labor, supply and waste were evaluated for HIV-1 and HCV bDNA on the semi-automated System 340 bDNA Analyzer and the automated VERSANT 440 Molecular System (V440). HIV-1 sample processing was evaluated using a 24- and 48-position centrifuge rotor. Vigilance time (hands-on manipulations plus incubation time except initial target hybridization) and disposables were approximately 37 and 12% lower for HIV-1, and 64 and 31% lower for HCV bDNA, respectively, with V440. Biohazardous solid waste was approximately twofold lower for both assays and other waste types were the same for either assay on both platforms. HIV-1 sample processing vigilance time for the 48-position rotor was reduced by 2 h. V440 provides cost savings and improved workflow.

