Related Experiment Video
Updated: Jul 10, 2026

Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Crosstalk triggered safety standby pacing associated with an improperly seated ventricular lead
Qi Sun1, Keping Chen, Shu Zhang
1Chinese Academy of Medical Sciences and Peking Union Medical College, Centre of Arrhythmia Diagnosis and Treatment, Fuwai Hospital, Beijing 100037, China.
Abstract:
Ventricular Safety Standby (VSS) was designed to prevent the potentially serious consequences of crosstalk in many dual-chamber pulse generators. In this report, we describe a patient in whom the VSS was activated in a special way. When the VSS presents during bipolar sensing configuration, the clinician should notice the proper electrical contact between leads and pulse generator.