Evaluation of the ventilator-user interface of 2 new advanced compact transport ventilators
François Templier1, Patrick Miroux, François Dolveck
1SAMU 92, Département d'Anesthésie Réanimation, Hôpital Raymond Poincaré, Assistance Publique Hôpitaux de Paris, 104 Boulevard Raymond Poincaré, 92380 Garches, France. francois.templier@rpc.aphp.fr
Background:
Mechanical ventilation during patient transport frequently utilizes compact portable pneumatic ventilators that have limited ventilator-settings options. New advanced transport ventilators should yield quality improvements, but their user-friendliness needs to be tested.
Objective:
To evaluate the ventilator-user interface of 2 new transport ventilators.
Methods:
This was a 2-center descriptive study in which the ventilator-user interfaces of the Oxylog 3000 and Elisée 250 were compared by 20 French senior emergency physicians who were initially unfamiliar with these ventilators. Each physician carried out 15 tasks with each ventilator and then assigned each ventilator a satisfaction score.
Results:
With the Elisée 250 the task success rate was significantly higher (85.6% vs 66.6% with the Oxylog 3000, p < 0.0001), and the total number of errors was lower (46 vs 113). The main errors were related to inspiratory flow settings with the Oxylog 3000 (31 errors), inspiratory-expiratory ratio settings with the Elisée 250 (11 errors), ventilation mode choice with the Oxylog 3000 (17 errors), trigger sensitivity setting with the Elisée 250 (16 errors) and the Oxylog 3000 (11 errors), and alarm range setting with the Oxylog 3000 (10 errors). The mean satisfaction score was significantly better with the Elisée 250 (81% +/- 7, range 64-92%) than with the Oxylog 3000 (66% +/- 10, range 49-87%) (p < 0.0001).
Conclusions:
The Elisée 250 ventilator-user interface was easier to use than that of the Oxylog 3000. The applicability of these results to other types of users will require further studies, but the types of errors found in our study might help future users.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: