Is hand injection of central venous catheters for contrast-enhanced CT safe in children?
Lane F Donnelly1, Julie Dickerson, John M Racadio
1Department of Radiology, Cincinnati Children's Hospital Medical Center, MLC 5031, 3333 Burnet Ave., Cincinnati, OH 45229-3039, USA. lane.donnelly@cchmc.org
Insights
Hand injecting contrast material via central venous catheters in children caused catheter damage in 0.3% of cases over six years. This highlights potential risks compared to power injection methods for CT scans.
Area of Science:
- Pediatric radiology
- Medical device safety
- Interventional radiology
Background:
- Central venous catheters are crucial for administering intravenous contrast material in pediatric CT scans.
- Hand injection is a common method, but its safety profile requires evaluation.
Purpose of the Study:
- To assess the safety and incidence of central venous catheter damage associated with hand injection of contrast material in pediatric patients.
Main Methods:
- Retrospective review of pediatric patients undergoing CT scans with contrast material administered via central venous catheters.
- Analysis of documented instances of central venous catheter damage during hand injection procedures.
Main Results:
- Four cases of central venous catheter damage occurred over a 6-year period during hand injection.
- This represents an incidence of 0.3% of all hand-injected central venous catheters during the study period.
Conclusions:
- Hand injection of contrast material poses a risk of central venous catheter damage in pediatric patients.
- The findings suggest that power injection may be a safer alternative, warranting consideration in institutional policies.
Objective:
Our objective is to review our safety experience with hand injection of central venous catheters for administration of i.v. contrast material for CT in children.
Conclusion:
Administration of i.v. contrast material by means of hand injection led to damage of four central venous catheters during a 6-year period at our institution (0.3% of central venous catheters hand injected during that time). In addition to the growing evidence of the safety of power injection of i.v. contrast material in central venous catheters, the potential danger of hand injection of central venous catheters should be considered when policies are made concerning the delivery of i.v. contrast material via central venous catheters.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies VII: Vascular Imaging
Cardiac Catheterization I: Pre-Procedure Overview
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...