Related Experiment Video
Updated: Jul 10, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Dose optimization for the MRI-accelerator: IMRT in the presence of a magnetic field
A J E Raaijmakers1, B Hårdemark, B W Raaymakers
1Department of Radiotherapy, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX, Utrecht, The Netherlands. a.raaymakers@umcutrecht.nl
Abstract:
A combined system of a 6 MV linear accelerator and a 1.5 T MRI scanner is currently being developed. In this system, the patient will be irradiated in the presence of a 1.5 T magnetic field. This causes a strong dose increase at tissue-air interfaces. Around air cavities in the patient, these effects may become problematic. Homogeneous dose distributions can be obtained around regularly shaped symmetrical cavities using opposing beams. However, for more irregularly shaped cavities this approach may not be sufficient. This study will investigate whether IMRT can be used to cope with magnetic field dose effects, in particular for target volumes adjacent to irregularly shaped air cavities. Therefore, an inverse treatment planning approach has been designed based on pre-calculated beamlet dose distribution kernels. Using this approach, optimized dose distributions were calculated for B = 1.5 T and for B = 0 T. Investigated target sites include a prostate cancer, a laryngeal cancer and an oropharyngeal cancer. Differences in the dose distribution between B = 0 and 1.5 T were minimal; only the skin dose increased for B = 1.5 T. Homogeneous dose distributions were obtained for target structures adjacent to air cavities without the use of opposing beams. These results show that a 1.5 T magnetic field does not compromise the ability to achieve desired dose distributions with IMRT.
More Related Videos
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Atomic Nuclei: Magnetic Resonance
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current