False-positive results on colorimetric carbon dioxide analysis in neonatal resuscitation: potential for serious
S M Hughes1, B L Blake, S L Woods
1Eudowood Neonatal Pulmonary Division, Department of Pediatrics, The Johns Hopkins University, Baltimore, MD, USA.
Abstract:
A term infant requiring resuscitation was found to have a false-positive color change on a colorimetric carbon dioxide device as a result of administration of epinephrine via an endotracheal tube. Using models of direct application and vapor exposure with a test lung, we discovered that epinephrine, atropine, infasurf and naloxone may result in false-positive color change. This false-positive response may lead to delayed recognition of esophageal intubation.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and HCO3− values are examined to...
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...
Acute Respiratory Failure-III
Acute Respiratory Failure-IV