Sensitivity of an eight-element phased array coil in 3 Tesla MR imaging: a basic analysis
Yoshiyasu Hiratsuka1, Hitoshi Miki, Keiichi Kikuchi
1Department of Radiology, Ehime University School of Medicine, Ehime, Japan. hirazka@m.ehime-u.ac.jp
Summary
The 8-element phased array head coil (8 ch coil) offers superior signal-to-noise ratio (SNR) and image uniformity for 3 Tesla MRI compared to conventional quadrature-type birdcage head coils (QD coils). This advancement leads to higher quality 3T MR images.
Area of Science:
- Magnetic Resonance Imaging (MRI)
- Medical Imaging Technology
Background:
- Conventional quadrature-type birdcage head coils (QD coils) are standard for 3 Tesla (3T) MRI.
- Evaluating advanced coil designs is crucial for improving image quality.
Purpose of the Study:
- To compare the performance of an 8-element phased array head coil (8 ch coil) against a QD coil.
- Assess signal-to-noise ratio (SNR) and image uniformity at 3T MRI.
Main Methods:
- Phantom studies using silicon oil were conducted on a 3T MR system.
- T(1)-weighted spin echo (SE) and T(2)-weighted fast SE images were acquired.
- Image quality was evaluated by SNR, uniformity, and visual assessment in healthy volunteers.
Main Results:
- The 8 ch coil demonstrated approximately double the SNR in the central phantom region compared to the QD coil.
- The 8 ch coil exhibited a flatter SNR distribution across the image, with less spatial variation than the QD coil.
- Improved image uniformity was observed with the 8 ch coil for both SE and fast SE sequences.
Conclusions:
- The 8-element phased array head coil significantly enhances SNR and image uniformity at 3T MRI.
- The 8 ch coil provides higher quality 3T MR images compared to conventional QD coils.
- This technology is beneficial for advanced 3T MR imaging applications.
Related Concept Videos
Magnetic Resonance Imaging
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
NMR Spectrometers: Resolution and Error Correction
When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
Magnetic Field Of A Current Loop
Consider a circular loop with a radius a, that carries a current I. The magnetic field due to the current at an arbitrary point P along the axis of the loop can be calculated using the Biot-Savart law.