Mechanical prosthetic valve dysfunction causing pulsus alternans leading to intermittent electromechanical
Vasha Kaur1, Martyn N N Robertson, Vipin Zamvar
1Royal Infirmary of Edinburgh, 51 Little France Crescent, Old Dalkeith Road, Edinburgh, Scotland EH16 4SA, UK.
Abstract:
We describe a case of electromechanical dissociation (EMD) in a 62-year-old female patient following insertion of a Medtronic Hall mitral valve prosthesis. She initially developed pulsus alternans which led, over a few hours, to intermittent electromechanical dissociation. Clinical and echocardiographic findings are described. Emergency surgical intervention revealed a piece of chorda wedged between the disc occluder and the valve ring. This case illustrates clinical and Doppler echocardiographic features associated with a rare presentation of an unusual perioperative complication of prosthetic mitral valve dysfunction. We have also included a brief review of related literature.
Related Concept Videos
Mitral Valve Prolapse I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Aortic Regurgitation I: Introduction
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...