Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
New circulating-water devices warm more quickly than forced-air in volunteers
Anupama Wadhwa1, Ryu Komatsu, Mukadder Orhan-Sungur
1Outcomes Research Institute, University of Louisville, Louisville, Kentucky, USA. anwadh01@louisville.edu
Background:
Newer circulating-water systems supply more heat than forced-air, mainly because the heat capacity of water is much greater than for that of dry warm air and, in part, because they provide posterior as well as anterior heating. Several heating systems are available, but three major ones have yet to be compared directly. We therefore compared two circulating-water systems with a forced-air system during simulation of upper abdominal or chest surgery in volunteers.
Methods:
Seven healthy volunteers participated on three separate study days. Each day, they were anesthetized and cooled to a core temperature near 34 degrees C, which was maintained for 45-60 min. They were then rewarmed with one of three warming systems until distal esophageal core temperature reached 36 degrees C or anesthesia had lasted 8 h. The warming systems were 1) energy transfer pads (two split torso pads and two universal pads; Kimberly Clark, Roswell, GA); 2) circulating-water garment (Allon MTRE 3365 for cardiac surgery, Akiva, Israel); and 3) lower body forced-air warming (Bair Hugger #525, #750 blower, Eden Prairie, MN). Data are presented as mean +/- sd; P < 0.05 was statistically significant.
Results:
The rate of increase of core temperature from 34 degrees C to 36 degrees C was 1.2 degrees C +/- 0.2 degrees C/h with the Kimberly Clark system, 0.9 degrees C +/- 0.2 degrees C/h with the Allon system, and 0.6 degrees C +/- 0.1 degrees C/h with the Bair Hugger (P = 0.002).
Conclusions:
The warming rate with the Kimberly Clark system was 25% faster than with the Allon system and twice as fast as with the Bair Hugger. Both circulating-water systems thus warmed hypothermic volunteers in significantly less time than the forced-air system.
More Related Videos
06:02A Contemporary Warming/Restraining Device for Efficient Tail Vein Injections in a Murine Fungal Sepsis Model
Published on: November 6, 2020
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
Related Concept Videos
Decreased Body Temperature
Mechanism of heat transfer
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Mechanisms of Heat Transfer II
Mechanisms of Heat Transfer I
Mechanisms of Heat Transfer
Conduction, accounting for approximately 3% of body heat loss at rest, is the process of exchanging heat between molecules of two materials in direct contact. This can result in both heat loss and gain. For instance, when the body is submerged in water, which conducts heat 20 times more effectively than air, it can either lose or gain significant heat.