Hemoximetry as the "gold standard"? Error assessment based on differences among identical blood gas analyzer devices
Hartmut Gehring1, Lutz Duembgen, Mareike Peterlein
1From the *Department of Anesthesiology, University Clinic of Schleswig-Holstein, Campus Luebeck, Germany; †Institute of Mathematical Statistics and Actuarial Science, University of Bern, Switzerland; ‡Institute of Clinical Chemistry, University Clinic of Schleswig-Holstein, Campus Luebeck, Germany.
Background:
The calibration and testing procedures of a pulse oximeter with arterial blood samples from healthy subjects are based on reference values from the hemoximeter. There are no tests to identify the accuracy of the reference devices. Because of this limitation and since the true values of oxygen saturation (sO2 in %) in blood samples were not known, we used the differences between two identical devices, A and B, for error assessment.
Methods:
Two identical devices, A and B, from five leading manufacturers were investigated. Seventy-two arterial blood samples from 12 healthy volunteers at three different levels of saturation between 100% and 70% sO2 were randomly evaluated by the test systems.
Results:
The observed differences (Delta) between Devices A and B, as a measure for the error of the hemoximeters, increased significantly with all manufacturers from level 97 (Deltamin, -0.9%; Deltamax, 2.6%) to 85 (Deltamin, -2.4%; Deltamax, 4.3), this effect was even stronger between levels 97 and 75 (Deltamin, -4.6%; Deltamax, 4.3%). A variance proportion analysis revealed the concentration of the reduced hemoglobin as the main error source for sO2 measurements. Independent from the sO2 levels there were also significant differences for the carboxy hemoglobin concentration in the range of 0%-4% and for the methemoglobin concentration in the range of 0%-1%.
Conclusions:
The variance of sO2 measurements between identical devices increased significantly when saturation decreased from the normal level of 97% to the hypoxemic levels of 85% and 75%.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...