Related Experiment Video
Updated: Jul 9, 2026

Rotating the Intraocular Lens to Prevent Posterior Capsular Opacification in Cataract Surgeries
Published on: July 7, 2023
Lightweight autoclavable wide-angle contact lens for vitreous surgery
K V Chalam1, Shailesh K Gupta, Swati Agarwal
1University of Florida College of Medicine, Department of Ophthalmology, 580 West 8th Street, Tower 2, 3rd Floor, Jacksonville, FL 32209, USA.
Abstract:
The authors describe an autoclavable, self-stabilizing, lightweight wide-angle contact lens for vitrectomy. The lens has two optical pieces with perforated plastic casing to sustain a high temperature (150 degrees C) for autoclaving. The lens has a 106 degrees static and 127 degrees dynamic field of view. The footplates and reduced weight (2.4 grams) due to the plastic casing allow self-stabilization of the lens. The open lens design.with high temperature resistant plastic prevents fogging during autoclaving and surgery. The autoclavable, self-stabilizing, lightweight wide-angle contact lens allows visualization of the peripheral retina during surgery and faster sterilization by autoclaving between surgeries without the disadvantage of lens fogging.
Related Concept Videos
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...