Multi-analyte Biochip (MAB) Based on All-solid-state Ion-selective Electrodes (ASSISE) for Physiological Research
Published on: April 18, 2013
Unbiased selectivity coefficients obtained for the pulsed chronopotentiometric polymeric membrane ion sensors
Hasini Perera1, Alexey Shvarev
1Department of Chemistry, Oregon State University, 153 Gilbert Hall, Corvallis, Oregon 97331-4003, USA.
Abstract:
We report here on the successful observation of the unbiased thermodynamic selectivity of ion-selective sensors working in normal pulse chronopotentiometric mode (pulstrodes). In contrast to ion-selective electrodes, the pulstrodes do not require careful counterbalancing of the transmembrane ionic fluxes to achieve unbiased thermodynamic selectivity. The pulstrodes can work under asymmetric conditions, which are often encountered in practice. The composition of the inner filling solution did not affect the sensor response, indicating that the transmembrane flux of primary ions was indeed effectively suppressed in the absence of ion exchanger. For the K-selective sensor considered here, an improvement of Mg discrimination by a factor of 1000 was demonstrated.
More Related Videos
Related Concept Videos
Potentiometry: Membrane Electrodes
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...