Comparison of biphasic and monophasic defibrillation waveforms in an isolated rabbit heart preparation
1Centre for Biomedical Technology, University of Technology Sydney, Australia.
Study Objective:
The aims were to develop a Langendorff rabbit heart model and to compare monophasic and biphasic defibrillation pulses.
Design:
Hearts were perfused with a Krebs-Henseleit solution and two 1.4 cm2 Pt-Ir mesh patch electrodes were sutured onto the ventricles. A 5 ms monophasic or 10 ms biphasic pulse, with randomly selected voltages of 30, 50, 70, 90, 110, or 130 V, defibrillated the heart after 10 s of fibrillation.
Subjects:
11 adult male New Zealand white rabbits weighing 2.8(0.27) kg, were used for the studies.
Measurements And Main Results:
A total of 72 fibrillation and defibrillation sequences were conducted in each preparation. The results were fitted to a sigmoidal dose-response curve by logistic regression analysis. Voltage and energy values from the fitted data at 50% and 80% success (V50, V80, E50, E80) indicated a significantly lower (p less than 0.05) defibrillation threshold voltage and energy for the biphasic waveform [V50 = 48 (SD19) V, V80 = 87(27) V, E50 = 0.15(0.12) J, E80 = 0.48(0.29) J] compared with the monophasic waveform [V50 = 79(20) V, V80 = 110(20) V, E50 = 0.27(0.12) J, E80 = 0.5(0.12) J]. There was no observed difference in defibrillation success rate between the first and second halves of any study.
Conclusions:
The Langendorff rabbit heart model is suitable for assessing electrical fibrillation and defibrillation mechanisms. Defibrillation can be achieved with a lower energy when using a biphasic rather than a monophasic pulse.