Related Experiment Video
Updated: Jul 9, 2026

Ultrafast Time-resolved Near-IR Stimulated Raman Measurements of Functional π-conjugate Systems
Published on: February 10, 2020
Sensitive H(2) detection by use of thermal-lens Raman spectroscopy without a tunable laser
Abstract:
A new nonlinear Raman spectroscopy technique for trace-gas detection was proposed and demonstrated. The technique involved the use of a thermal-lens detection scheme to monitor thermal emission from the stimulated Raman process. We termed this technique thermal-lens Raman spectroscopy, and it was combined with a novel scheme involving a nonlinear Raman spectroscopy without a tunable laser. This technique was applied to detecting trace hydrogen molecules in the atmosphere by use of a pulsed Nd:YAG laser and a continuous-wave He-Ne probe laser. A detection limit of 9 parts in 10(6) was attained.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...