[Pineapple juice as a negative oral contrast agent in magnetic resonance cholangiopancreatography]
Related Experiment Video
Updated: Jul 9, 2026

Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging
Published on: April 4, 2013
1Service de radiologie, Assistance-Publique-Hôpitaux de Paris, Faculté de Médecine Pierre et Marie Curie, Hôpital Saint-Antoine, 184, rue du Faubourg Saint-Antoine, 75012 Paris. lionel.arrive@sat.ap-hop-paris.fr
Purpose:
The quality of magnetic resonance cholangiopancreatography (MRCP) images is frequently degraded by high signal from the gastrointestinal tract on heavily T2W images. The purpose of this study is to evaluate pineapple juice (PJ) as an oral negative contrast agent in MRCP.
Materials And Methods:
Results from MRCP in 50 patients with PJ and 50 patients with paramagnetic contrast (ferumoxsil-Lumirem) were compared. Reviewers were blinded to the type of contrast agent. Exam quality was recorded with regards to signal suppression in the stomach, duodenum and proximal small bowel and with regards to pancreatic duct and biliary ducts visualization. In vitro, the signal characteristics of several commercially available brands of PJ were assessed using T1W, T2W and MRCP sequences. Signal intensity was correlated with the manganese concentration measured using atomic absorption spectrometry. Finally, the reviewers compared the taste of PJ and ferumoxsil.
Results:
On MRCP sequences, results were similar with regards to signal suppression in the stomach, duodenum and proximal small bowel with PJ and ferumoxsil. Visualization of the pancreatic duct, intrahgepatic bile ducts and CBD was similar with PJ and ferumoxsil. The signal intensity of commercially available brands of PJ on T2W and MRCP sequences correlated well with the measured manganese concentration on spectroscopy. Variations in manganese concentration were observed, with values ranging from 3.65 to 27.24 mg/L. The reviewers noted that PJ tasted "good" or "very good" and that ferumoxsil tasted "bad" or "very bad".
Conclusion:
Ingestion of PJ provides effective signal suppression in the GI tract on MRCP, similar to paramagnetic contrast agents. Because manganese concentration is highly variable in commercially available PJ brands, a brand with high manganese concentration should be selected.
Insights
Pineapple juice effectively suppresses gastrointestinal signal on magnetic resonance cholangiopancreatography (MRCP) scans, comparable to traditional agents. Choose brands with high manganese for optimal results in MRCP imaging.
Area of Science:
- Radiology
- Medical Imaging
- Gastroenterology
Background:
- Magnetic resonance cholangiopancreatography (MRCP) image quality is often compromised by high gastrointestinal signal on T2-weighted images.
- Effective oral negative contrast agents are needed to improve MRCP diagnostic accuracy.
Purpose of the Study:
- To evaluate pineapple juice (PJ) as a potential oral negative contrast agent for MRCP.
- To compare the efficacy of PJ with a standard paramagnetic contrast agent (ferumoxsil).
Main Methods:
- MRCPs were performed on 100 patients (50 with PJ, 50 with ferumoxsil).
- Image quality, including signal suppression and ductal visualization, was assessed by blinded reviewers.
- In vitro analysis correlated PJ signal intensity with manganese concentration.
- Taste comparison between PJ and ferumoxsil was conducted.
Main Results:
- Pineapple juice demonstrated comparable signal suppression in the stomach, duodenum, and small bowel to ferumoxsil.
- Visualization of the pancreatic and biliary ducts was similar between the two agents.
- PJ's signal intensity correlated with manganese concentration, which varied significantly among brands.
- Pineapple juice was preferred for taste over ferumoxsil.
Conclusions:
- Pineapple juice serves as an effective negative oral contrast agent for MRCP, yielding results similar to paramagnetic agents.
- Selection of PJ brands with high manganese concentration is recommended for optimal performance.
- The palatable taste of PJ offers a patient-friendly alternative for MRCP contrast enhancement.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging
Endoscopic Procedures V: ERCP
Patient...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies III: Computed Tomography
Imaging Studies IV: Magnetic Resonance Imaging