Thin-section CT of lung without ECG gating: 64-detector row CT can markedly reduce cardiac motion artifact which can
Masahiro Yanagawa1, Noriyuki Tomiyama, Hiromitsu Sumikawa
1Department of Radiology, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita-city, Osaka 565-0871, Japan. m-yanagawa@radiol.med.osaka-u.ac.jp
Purpose:
Motion artifacts, which can mimic thickened bronchial wall and the cystic appearance of bronchiectasis, constitute a potential pitfall in the diagnosis of interstitial or bronchial disease. Therefore, purpose of our study was to evaluate whether 64-detector row CT (64-MDCT) enables a reduction in respiratory or cardiac motion artifacts in the lung area on thin-section CT without ECG gating, and to examine the correlation between cardiac motion artifact and heart rate.
Materials And Methods:
Thirty-two patients with suspected diffuse lung disease, who underwent both 8- and 64-MDCT (gantry rotation time, 0.5 and 0.4s, respectively), were included. The heart rates of an additional 155 patients were measured (range, 48-126 beats per minute; mean, 76 beats per minute) immediately prior to 64-MDCT, and compared to the degree of cardiac motion artifact. Two independent observers evaluated the following artifacts on a monitor without the knowledge of relevant clinical information: (1) artifacts on 8- and 64-MDCT images with 1.25-mm thickness and those on 64-MDCT images with 0.625-mm thickness in 32 patients; and (2) artifacts on 64-MDCT images with 0.625-mm thickness in 155 patients.
Results:
Interobserver agreement was good in evaluating artifacts on 8-MDCT images with 1.25-mm thickness (weighted Kappa test, kappa=0.61-0.71), and fair or poor in the other evaluations (kappa<0.31). Two observers stated that cardiac motion artifacts were more significant on 8-MDCT than on 64-MDCT in all 32 patients. Statistically significant differences were found at various checkpoints only in comparing artifacts between 8- and 64-MDCT for 1.25-mm thickness (Wilcoxon's signed-rank test, p<0.0017). Cardiac motion artifacts on 64-MDCT had no significant correlation with heart rate (Spearman's correlation coefficient by rank test).
Conclusion:
The high temporal resolution of 64-MDCT appears to reduce cardiac motion artifact that can affect thin-section scans of the lung parenchyma.
More Related Videos
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
08:05Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia
Published on: December 19, 2020
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...