Development and validation of a noninvasive method to estimate cardiac output using cuff sphygmomanometry
Shiu-Shin Chio1, Jeffrey J Tsai, Yen-Ming Hsu
1Pulse Metric, Inc.
Background:
Obtaining cardiac output (CO) measurements noninvasively during routine blood pressure recording can improve hypertension management. A new method has been developed that estimates cardiac output using pulse-waveform analysis (PWA) from a brachial cuff sphygmomanometer. This study evaluates the ability of PWA to track changes in CO as derived by Doppler ultrasound during dobutamine stimulation.
Hypothesis:
This study aims to validate the PWA CO estimation over a wide CO range as would be obtained by dobutamine stimulation during Doppler ultrasound evaluation.
Method:
A total of 48 patients undergoing standard dobutamine stress echocardiography testing for accepted clinical indications were enrolled. Among them, 44 patients (age 36-83, 18 females, 26 males) with good waveform data for analyses provided estimates of CO in this study. Noninvasive measurements of CO were performed using both Doppler ultrasound recordings and PWA techniques simultaneously at each stage of dobutamine infusion.
Results:
A total of 207 simultaneous pulse-waveform analyses and Doppler measurements were taken during dobutamine stress on 44 cardiac patients. Linear regression analysis revealed good intra-patient correlation between pulse-waveform analysis and Doppler at different dobutamine-induced CO with coefficients from r = 0.69 to 0.98 (p < 0.05). Analysis of all patients yielded an overall correlation of r = 0.82 (p < 0.001, bias = 0.4 L/min, standard deviation = 1.8 L/min).
Conclusion:
The CO measured noninvasively from a sphygmomanometer using this PWA method correlates well with those of Doppler through a range of dobutamine-stimulated levels. The CO by PWA should be useful for monitoring hemodynamic changes in hypertensive and cardiac patients during routine blood pressure measurement.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: