Use of an intracardiac electrogram eliminates the need for a surface ECG during implantable
Kevin A Michael1, Brett J Peterson, Arthur M Yue
1Wessex Cardiothoracic Centre, Southampton University Hospital Trusts, Southampton, UK.
Background:
A surface electrocardiogram (SECG) for pacing threshold measurements during routine implantable cardioverter-defibrillator (ICD) follow-up can be cumbersome. This study evaluated the use of an intrathoracic far-field electrogram (EGM) derived between the Can and superior vena cava (SVC) electrode -- the Leadless electrocardiogram (LLECG), in dual chamber ICDs in performing pacing threshold tests.
Methods:
The LLECG was evaluated prospectively during atrial and ventricular pacing threshold testing as a substudy of the Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators trial (EMPIRIC) in which dual chamber ICDs were implanted in 888 patients. Threshold tests were conducted at 1 volt by decrementing the pulse width. Follow-up at three months compared pacing thresholds measured using LLECG with those using Lead I of the surface ECG (SECG). The timesaving afforded by LLECG was assessed by a questionnaire.
Results:
The median threshold difference between LLECG and SECG measurements for both atrial (0.00 ms, P = 0.90) and ventricular (0.00 ms, P = 0.34) threshold tests were not significant. Ninety percent of atrial and ventricular threshold differences were bounded by +/- 0.10 ms and -0.10 to +0.04 ms, respectively. We found that 99% of atrial and ventricular thresholds tests at six and 12 months attempted using LLECG were successfully completed. The questionnaire indicated that 65% of healthcare professionals found LLECG to afford at least some timesaving during device follow-ups.
Conclusion:
Routine follow-up can be performed reliably and expeditiously in dual chamber Medtronic (Minneapolis, MN, USA) ICDs using LLECG alone, resulting in overall timesaving.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials