Accelerometer-derived time intervals during various pacing modes in patients with biventricular pacemakers:
Frank I Marcus1, Vincent Sorrell, John Zanetti
1University Medical Center, Sarver Heart Center, Tucson, AZ 85724, USA. fmarcus@u.arizona.edu
Introduction:
Changes due to biventricular pacing have been documented by shortening of QRS duration and echocardiography. Compared to normal ventricular activation, the presence of left bundle branch block (LBBB) results in a significant change in cardiac cycle time intervals. Some of these have been used to quantify the underlying cardiac dyssynchrony, assess the effects of biventricular pacing, and guide programming of ventricular pacing devices. This study evaluates a simple noninvasive method using accelerometers attached to the skin to measure cardiac time intervals in biventricularly paced patients.
Methods:
Ten patients with biventricular pacemakers previously implanted for congestive heart failure were paced in the AAI mode, then in atrioventricular (AV) sequential mode from the right and left ventricles followed by biventricular pacing. Simultaneous recordings were obtained by 2D, Doppler echocardiography as well as by accelerometers. Similar recordings were obtained from 10 gender, aged matched, normal controls during sinus rhythm.
Results:
Compared to normals, heart failure patients paced in AAI mode had prolonged isovolumetric contraction time (IVCT), shorter ventricular ejection time (LVET), and prolonged isovolumetric relaxation (IVRT). With biventricular pacing the IVCT decreased, but the LVET and IVRT did not change significantly. There was excellent correlation between the echo and accelerometer-measured intervals.
Conclusions:
Shortening of the IVCT measured by an accelerometer is a consistent time interval change due to biventricular pacing that probably reflects more rapid acceleration of left ventricular ejection. The accelerometer may be useful to assess immediate efficacy of biventricular pacing during device implantation and optimize programmable time intervals such as AV and interventricular (VV) delays.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Electrophysiology of Normal Cardiac Rhythm