Determination of urethral catheter surface lubricity
Katarzyna Kazmierska1, Maciej Szwast, Tomasz Ciach
1Faculty of Chemical and Process Engineering, Warsaw University of Technology, ul. Warynskiego 1, Warszawa, 00-645, Poland. kazmierska@ichip.pw.edu.pl
Abstract:
Device for in-vitro measurement of static and kinetic friction coefficient of catheter surface was developed. Tribometer was designed and constructed to work with exchangeable counter-faces (polymers, tissue) and various types of tubes, in wet conditions in order to mimic in-vivo process. Thus seven commercially available urethral catheters, made from vinyl polymers, natural latex with silicone coating, all-silicone or hydrogel coated, and one made from polyvinylchloride with polyurethane/polyvinylpyrrolidone hydrogel coating obtained in our laboratory, were tested against three various counter faces: polymethacrylate (organic glass), inner part of porcine aorta and porcine bladder mucosa. Additionally, the hydrophility/hydrophobity of tested catheters was stated via water wetting contact angle measurement. Super-hydrophilic biomaterials revealed low friction on tissue and hydrophobic counter-face; slightly hydrophobic showed higher friction in both cases, while more hydrophobic manifested low friction on tissue but high on hydrophobic polymer. The smoothest friction characteristic was achieved in all cases on tissue counter-faces. The measured values of the static coefficient of friction of catheters on bladder mucosa counter-face were as follows: the highest (0.15) for vinyl and siliconised latex catheters and 3 folds lower (0.05) for all-silicone ones. Hydrogel coated catheters exhibited the lowest static and kinetic friction factors.
Related Concept Videos
Fineness of Cement
Direct...
Design Example: Deciding Thickness of Lubricating Fluid in a Shaft
To calculate the required thickness of the lubricant layer, the tangential velocity at the shaft's surface must first be determined. This velocity is calculated by converting the rotational speed to angular velocity...