Abstract:
We introduce an imaging technique that combines complementary features of ultrasound and near-infrared diffusive light imaging. We achieve the combined technology experimentally by mounting an ultrasound array together with multiple laser source and optical detector fibers upon a hand-held probe. The technique is demonstrated with tissue phantoms wherein both acoustic and optical sensors image the volume underneath the probe. Coregistration of acoustic and optical images is achieved with an accuracy of 0.27+/-0.20cm, approximately half of the image pixel size of our prototype. Accurate determination of target optical absorption is also achieved by use of image segmentation on the ultrasound reconstruction. The combined technique may provide improved breast-cancer detection sensitivity and specificity.