Central venous catheter infusions: a laboratory model shows large differences in drug delivery dynamics related to
Mark A Lovich1, Gregory L Peterfreund, Nathaniel M Sims
1Department of Anesthesia and Pain Medicine, Caritas St. Elizabeth's Medical Center, Brighton, MA, USA.
Objective:
Central venous catheters (CVCs) are conduits for drug infusions. Dead volumes of different CVC lumens vary considerably. This study quantitatively evaluated drug delivery dynamics of CVCs in a laboratory model of continuous drug infusion.
Design:
CVCs studied included a triple-lumen catheter (16-gauge and 18-gauge lumens), the proximal infusion port of a pulmonary artery catheter, and a 9-Fr introducer sheath, with and without a pulmonary artery catheter in the lumen. One syringe pump infused a carrier. A second pump infused the model drug methylene blue (3 mL/hr), joining the carrier immediately upstream of the CVC. Samples were collected every minute for quantitative analysis.
Setting:
Laboratory model.
Subjects:
None.
Interventions:
At low fixed flow rates, experiments characterized drug delivery kinetics of different CVCs. Data collection then assessed effects of increased carrier flow.
Measurement And Main Results:
The time to steady-state delivery after initiation of methylene blue infusion differed between CVCs. At a carrier flow of 10 mL/hr, the fastest achievement of steady-state delivery was with the 18-gauge lumen of a triple-lumen catheter. The 9-Fr introducer had the slowest time to achieve steady-state delivery. Other CVCs had intermediate kinetics. Reducing drug delivery from steady state to zero after cessation of methylene blue infusion was fastest with the 18-gauge lumen and slowest with the 9-Fr introducer. Increasing carrier flow rates from 10 to 60 mL/hr hastened the time to target for initiation and cessation of methylene blue delivery.
Conclusions:
Experiments demonstrate large differences between CVCs in the dynamics for delivery of model drug methylene blue. Achieving targeted steady-state delivery, and termination of a planned continuous drug infusion, may be far slower than typically appreciated. Delivery kinetics depend on the dead volume and the rate of carrier flow. Safe and effective management of continuous drug infusions depends on understanding the dynamics of the delivery system.
Related Concept Videos
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Compartment Models: Single-Compartment Model
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
Compartment Models: Two-Compartment Model
One-Compartment Open Model for IV Bolus Administration: Estimation of Elimination Rate Constant, Half-Life and Volume of Distribution