Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Transcutaneous PCO2 monitoring during initiation of noninvasive ventilation
Jan H Storre1, Boris Steurer, Hans-Joachim Kabitz
1Department of Pneumology, University Hospital Freiburg, Killianstrasse 5, D-79106 Freiburg, Germany.
Background:
To assess the efficacy of transcutaneous Pco2 (Ptcco2) measurements for monitoring alveolar ventilation in patients requiring noninvasive positive-pressure ventilation (NPPV).
Methods:
In a prospective study on method agreement pairs of Paco2 and Ptcco2 (SenTec Digital Monitor; SenTec AG; Therwil, Switzerland), measurements were performed every 10 min during the establishment of NPPV over a 4-h period in 10 patients (8 patients with COPD) presenting with acute-on-chronic hypercapnic respiratory failure, thus providing 250 pairs of measurement.
Results:
Mean (+/- SD) Paco2 decreased from 67.2 +/- 11.9 mm Hg (Ptcco2, 65.5 +/- 13.9 mm Hg) to 54.6 +/- 8.8 mm Hg (Ptcco2, 47.8 +/- 8.8 mm Hg), and mean pH increased from 7.36 +/- 0.03 to 7.44 +/- 0.04. Following Ptcco2 assessment, Ptcco2 in the ensuing 2-min period was the strongest predictor for Paco2 compared to Ptcco2 in the ensuing 5-min period and to real-time measurements. Ptcco2 was highly correlated with Paco2 (r = 0.916; p < 0.001), as determined by linear regression analysis. The mean difference between Paco2 and Ptcco2 was 4.6 mm Hg, and the limits of agreement (bias +/- 1.96 SDs) ranged from -3.9 to 13.2 mm Hg, following the Bland and Altman analysis. Retrospective drift correction produced an even higher correlation (r = 0.956; p < 0.001) with lower limits of agreement (-1.7 to 7.5 mm Hg).
Conclusions:
Ptcco2 measurements provide a sensitive, continuous, and noninvasive method for monitoring alveolar ventilation in patients who are receiving short-term NPPV therapy. Drift correction of Ptcco2 measurements improves the accuracy of Ptcco2 monitoring compared to the "gold standard" Paco2 assessment. A lag time of approximately 2 min is present for reliable Ptcco2 values compared to Paco2 values. However, individual variance between Paco2 and Ptcco2 cannot be excluded.
Trial Registration:
www.uniklinik-freiburg.de/zks/live/uklregister/Oeffentlich.html Identifier:UKF001271.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Mechanical Ventilation I: Indication and Settings